Inaccuracy of routine creatinine measurement in canine urineCatherine Trumel, Armelle Dique´ lou, Herve´ Lefebvre, Jean-Pierre Braun http://www.woodleyequipment.com/documents/erd_article.pdf |
|
| Table 1. Comparison of the 5 methods
tested for canine urine creatinine measurement (manufacturers’ information). |
|||
| Instrument |
Method |
Temperature |
Analytical Range (mg/L) |
| Elimat (Jaffe´ ) |
Kinetic (10–120
s) Calibration by 20 mg/L solution |
37C |
1.7–150 |
| Elimat (Enzymatic) |
Creatininase Calibration by 20 mg/L solution |
37C |
1.4–200 |
| Reflovet Creatininase |
Reagent strip Calibration by magnetic strip |
37C |
5–100 |
| Vettest Creatininase |
Multilayer slide Calibration by bar code |
37C |
0.5–140 |
| Vitros Creatininase |
Multilayer slide Calibration by 5.2, 15.6, and 125.6 mg/L solutions |
37C |
0.5–140 |
| Table 2. Comparison of the results of urine (U)-creatinine measurements by 5 different techniques in 104 canine urine samples. | ||||||
| U-Creatinine (mg/L) |
||||||
| Elimat |
Reflovet |
Vettest |
Vitros |
|||
|
||||||
| Mean |
|
1204 |
991 |
1029 |
||
| Median |
|
1078 |
858 |
922 |
||
| SD |
|
776 |
602 |
626 |
||
| Minimum |
|
129 |
118 |
131 |
||
| Maximum |
|
3384 |
2717 |
2857 |
||
|
| Figure 1.
Scatterplot of results of U-creatinine measurements in 104 canine urine samples
using the Jaffe´ reaction assay and an enzymatic technique on the Elimat
analyzer (solid line = equivalence; dotted line = regression). |
|
|
| Figure 2. Scatterplot
(left) of results of U-creatinine measurements of 104 canine urine samples
using the Reflovet Plus and the mean of results obtained using the Jaffe´
reaction and an enzymatic technique with the Elimat (black line = equivalence).
Difference plot (right) of results according to the mean of results using
the Elimat analyzer. |
|
|
| Figure 3. Scatterplot
(left) of results of U-creatinine measurements in 104 canine urine samples
using the Vettest and the mean of results obtained using the Jaffe´
reaction and an enzymatic technique with the Elimat (black line = equivalence);
Difference plot (right) of results according to the mean of results using
the Elimat analyzer. |
|
|
| Figure 4. Scatterplot
(left) of results of U-creatinine measurements in 104 canine urine samples
using the Vitros and the mean of results obtained using the Jaffe´
reaction and an enzymatic technique with the Elimat (black line = equivalence);
Difference plot (right) of results according to the mean of results using
the Elimat analyzer. |